AGILENT E3640A PDF

Agilent. EA – EA. Programmable DC Power Supplies. Data Sheet. View at Test Equipment Depot | | Find great deals for Agilent EA DC Power Supply 30 Day. Shop with confidence on eBay!. On Sale Agilent EA. We take purchase orders and can provide special school, government and corporate pricing.

Author: Shanris Meshura
Country: Armenia
Language: English (Spanish)
Genre: Music
Published (Last): 17 November 2014
Pages: 129
PDF File Size: 11.86 Mb
ePub File Size: 16.98 Mb
ISBN: 955-3-70467-343-9
Downloads: 82485
Price: Free* [*Free Regsitration Required]
Uploader: Akinozragore

The front panel is easily separated from the main body of the power supply.

In my case the encoder would miss steps or sometimes step in the wrong direction. Accredited Calibration with Data.

A couple days later I had a new agiilent encoder ready to be installed. Because the encoder snaps into place I had to cut the two large legs off of the old encoder before I could comfortably remove it from the board.

Transcat makes it easy: Fortunately Agilent has an online shop that supplies many wear-and-tear items for their various products. Accredited Calibration without Data. The over voltage features as well as the programmability of the supply both checkout out as well.

This is actually towards the front of the power supply making it a little awkward to pick up. Request a Calibration Quote. By making it simple to connect, control instruments, and automate test sequences you can quickly move past the test development phase and access results faster than ever before with just a few clicks. It is also the same encoder used in many different Agilent products.

Related Posts  RSI THE COMPLETE GUIDE BY JOHN HAYDEN PDF

I recently bought a used Agilent power supply in a very sad condition. Ships in weeks. This lead me to believe that the only issue with this power supply was that the rotary encoder was on its last legs.

E3640A 30W Power Supply, 8V, 3A or 20V, 1.5A

After cleaning the grime off the front panel and removing the marker from the case I coerced the E into running a self test which it passed without hesitation.

Other than that there are a lot of passive components and a MOV soldered on near the output. Finally the board is out and the rotary encoder can be replaced! After doing a little research I found that it was a known issue for the rotary encoders in these power supplies to fail. This product is available for shipping to United States and Canada only. Now to actually get it installed. Don’t forget the Software! Request a Rental Quote. Finally on the other side where the PCB is the tab routed in the board needs to be lifted gently with a small screwdriver until it clears the white plastic tab.

Rent From Transcat Do you need a test instrument for a short time?

Keysight Agilent EA DC Power Supply | Transcat

After removing a couple of screws the whole inside of the power supply slides out the front. I made sure the holes had minimal solder before pressing the new encoder into place and soldering. Would you like this item calibrated?

Related Posts  OIME BIEN SATANAS PDF

Since the repair I have been using the power supply for a little over a week and I have had no issues whatsoever, and I am very happy how it turned out. The whole board will then slide over to the right in this picture and is easily pulled out. Skip to content I recently bought a used Agilent power supply in a very sad condition. Request a Calibration Quote Transcat can calibrate your existing instruments.

Select a Web Site

Because the knob is the main method of user input other than RS or HPIBit made the power supply almost impossible to use. I cleaned up the flux left over with a little rubbing alcohol and put everything back together.

It does have two speeds loud-ish and louder? You must have JavaScript enabled in your browser to utilize the functionality of this website. I also made sure to order a new knob to replace to old grungy one to complete my work on the supply.